Title: | THE DESIGN OF A SIGNAL CONDITIONING & ACQUISITION ELEMENTS OF A CHOPPED BROADBAND RADIATION PYROMETER |
Authors: | Jack, Kufre E. Etu, Israel A. Ukanide, Veronica N. |
Keywords: | Broadband Radiation Pyrometer Temperature Sensor Instrument Chopped Detector |
Issue Date: | Apr-2015 |
Publisher: | IJRET: International Journal of Research in Engineering and Technology |
Citation: | Kufre Esenowo Jack, Israel Agwu Etu & Veronica Nicholas Ukanide (2015). The design of a signal conditioning and acquisition elements of a chopped broadband radiation pyrometer that is capable of measuring temperature between 900oC and 1200oC, Published in International Journal of Research in Engineering and Technology (IJRET) Vol.4, Issue 4, 2015.pp. 461-470. |
Abstract: | This paper presents the design of signal conditioning and acquisition elements of a chopped broadband radiation pyrometer. This instrument is capable of measuring temperature between 900oC and 1200oC. This work aims at solving the problem of measuring hot objects with a thermometer. The radiation pyrometer is a non-contact temperature sensor that infers the temperature of an object by detecting its naturally emitted thermal radiation. It collects the visible and infrared energy and focuses it on a detector. The detector used in this device is a thermal sensor. It receives heat energy reflected from a mirror inclined at 45o to the incident signal from the hot object. The design achieved the following: temperature range measured, from 900℃ to 1200℃; the calibrated instrument is fairly linear with a tolerable non-linearity of 3.6%, with the sensitivity of 0.014𝑉℃−1 ; the resolution was quite very small as such it can easily detect the slightest change at its input; the rotating shutter was configured to supply the chopped signal; it operates at a frequency of 50Hz that is lower than the system frequency of 200Hz; the data acquisition system was able to capture data at a periodic time of 0.02 second and below, the system operates within the specified sampling range thus, satisfying Nyquist criteria. The signal so received by the detector is translated to a human readable form and sent to a display. |
URI: | http://repository.futminna.edu.ng:8080/jspui/handle/123456789/10297 |
Appears in Collections: | Mechatronics Engineering |
Files in This Item:
File | Description | Size | Format | |
---|---|---|---|---|
IJRET20150404082-THE DESIGN OF A SIGNAL CONDITIONING & ACQUISITION.pdf | 1.07 MB | Adobe PDF | View/Open |
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.